Modular gas microchromatograph. The instrument features two independent GC modules, each with an injector, WCOT and PLOT silica columns, and its own microelectromechanical system (MEMS) thermal conductivity detector (TCD), offering a detection limit of 1 ppm. It detects gases up to C12. The instrument performs rapid analysis: 1–3 minutes. It detects gases in the gas phase up to C12.
  • Maximum sample pressure 1000 psi. Recommended minimum pressure 10 psi.
  • Sample volume variable between 1 and 30 μL.
  • Maximum heating temperature 250°C.

Applications

  • Online chemical analysis (characterization) of compounds in gas streams (typically natural gas).
  • Q-Bond column for high-boiling-point hydrocarbons (up to C-7) and Rxi-1ms column specifically designed for VOCs.
  • Conventional and expanded natural gas analysis.
  • Dissolved Gas Analysis (DGA).
  • Monitoring of flue gases and stack emissions.